- Used Night Shift with the Celeste+ App
- Beginning in August 2025 the Celeste+ app will enable:
- Use of either the neck strap or chest belt for position information.
- Once configured for the Celeste+ App, a Night Shift device CANNOT:
- Be used by patients for long-term therapy.
- Interface with the Night Shift portal or Night Shift app to generate compliance reports.
- Functionality
- A fully charged Night Shift can record up to 3-nights and 12 hours per night with the Celeste+ App.
- The Night Shift will indicate the number of nights (3, 2, or 1 night) it can provide positional information before needing to be recharged
- When it is turned on, the Night Shift will vibrate 3, 2 or 1 time.
- The green LED will blink 3, 2 or 1 time for the first 5-minutes and then be a steady dimmed green.
- Night Shift LED blink patterns:
- When connected to the Celeste+ app:
- Bright green
- Dimmed green 5 minutes after the start of acquisition
- When it needs to be recharged:
- Amber blink every second plus vibration once every 5 seconds
- While charging:
- Steady green
- Off β when charging has completed
- The Night Shift can be recharged in 2 hours with USB C connector.
- Site Selection β Neck vs. Chest
- Detection of the supine position causing the airway to collapse is best measured at the neck.
- The accuracy of detected supine position versus video confirmation of the supine position is Neck r = 0.93, chest r = 0.78 (see Figure 3).
- 40% of participants reported needing to move their chest strap back to the correct position.
- 6.5% of participants were identified with supine positional OSA using a head-based sensor that were not detected using a trunk-based sensor.
- Using the Neck Strap
- Adjusting the Neck Strap
- This video explains how to adjust the Night Shift neck strap.
- Using the Chest Belt
- Adjusting the Chest Belt
- Affix the Night Shift to the Chest Belt.
- Use the Celeste+ App to change the setting from Neck Strap to Chest Belt.
- Adjust the Chest Belt so it remains centered during the night. The Chest Belt is designed to be worn over a shirt or other nightwear.
- Attach the Chest Belt Extension when needed, i.e., the chest measurement where the belt is worn is >48 inches. The Chest Belt Extension enables the belt to be worn by those with chest sizes up to 68 inches.
- Cleaning
- Detach the neck straps / chest belt from the device.
- Night Shift Device
- Wipe all surface areas of the device enclosure with a 70% isopropyl alcohol wipe. All surface areas should remain wet for at least 15 seconds.
- Using a new alcohol wipe, repeat the cleaning; allow to remain wet for 15 seconds. If any visible soil remains, repeat as needed. Allow to air-dry.
- Neck Strap / Chest Belt
- Wash each strap / chest belt separately by submerging in solution of 1 teaspoon (5 ml) of mild dish detergent (e.g. Dawn) per gallon (4 liters) of water.
- Agitate slightly for one to two minutes.
- Rinse with clear warm tap water for one minute.
- Dry with a clean paper towel. Ensure the straps / chest belt are completely dry before use.
- Warranty
- Use this registration page to register each device
- Neck strap β 6-months
- Common failure not covered by warranty β pulling on the magnetic clasps to release, causing the silicone to fail
- Chest strap β 6 months
- Typical useful life >12 months
- Battery β 12-months
- Typical useful life 3+ years
- Identified when recharged device wonβt support 3 nights with typical feedback time
- Failure rates increase if device sits for > 3 months without charging
- PCB electronic components within enclosure β 24-months
- For warranty requests, contact NightShift@advanced-sleep.com with serial number of device and details of malfunction.
- Integration and Study Results
- Please contact support@ensodata.com.