Night Shift Device Test

After being turned on, the Night Shift is designed to wait 15 or 30 minutes before delivering positional vibration. To skip the delay and test the vibrations on the device, follow these steps:

Step 1: Charge your device

Fully charge the device using a known good cable and power source. When the charge is complete, the green LED will turn off. Unplug the device from the charger.

Step 2: Test your Night Shift’s vibration

Please follow the steps below to test your Night Shift’s vibration motors.

  1. Unplug the device from any power source.
  2. Put the device into “test” mode by holding the on/off button down continuously (about 7-seconds) until the device vibrates.
  3. If you are using the Neck Strap, with the blue label facing down, place the device on a flat surface. Place two fingers on the black label of the device to feel the vibrations. If you are using the Chest Belt, with the blue label facing up, place the device on a flat surface. Place two fingers on the blue label of the device to feel the vibrations.
  4. After 30 seconds, the device should begin vibrating. After the first six vibrations, did the intensity increase? Wait another six vibrations, did it increase again? There should seven levels of intensity with six vibrations at each level. Following the six vibrations at the seventh level, the vibration intensity and frequency should begin a random pattern.

If the problem is replicated after following the steps above, click the following button to submit a support ticket. Warranty information for your Night Shift can be found here: https://www.advancedbrainmonitoring.com/warranty *For outside of North America, please contact your distributor with any technical support questions.

Support Ticket

© 2014-2024 Advanced Brain Monitoring, Inc.